Sign in →

Test Code CD40 B-Cell CD40 Expression by Flow Cytometry, Blood

Reporting Name

CD40 by Flow, QL, B

Useful For

Evaluating patients for hyper-IgM type 3 (HIGM3) syndrome due to defects in CD40, typically seen in patients less than 10 years

 

Assessing B-cell immune competence in other clinical contexts, including autoimmunity, malignancy, and transplantation

Performing Laboratory

Mayo Clinic Laboratories in Rochester

Specimen Type

Whole Blood EDTA


Ordering Guidance


This test is not used to detect in CD40L expression (CD154), which is responsible for X-linked hyper-IgM syndrome (HIGM1); see XHIM / X-Linked Hyper IgM Syndrome, Blood.



Shipping Instructions


Testing performed Monday through Friday. Specimens not received by 4pm (CST) on Friday may be cancelled.

 

Collect and package specimen as close to shipping time as possible. It is recommended that specimens arrive within 24 hours of collection.

 

Samples arriving on the weekend and observed holidays may be canceled.



Necessary Information


Ordering healthcare professional name and phone number are required.



Specimen Required


Container/Tube: Lavender top (EDTA)

Specimen Volume: 3 mL

Collection Instructions: Send whole blood specimen in original tube. Do not aliquot.

Additional Information: For serial monitoring, it is recommended that specimens are collected at the same time of day.


Specimen Minimum Volume

1 mL

Specimen Stability Information

Specimen Type Temperature Time Special Container
Whole Blood EDTA Ambient 72 hours PURPLE OR PINK TOP/EDTA

Reference Values

Present (normal)

Day(s) Performed

Monday through Friday

Test Classification

This test was developed using an analyte specific reagent. Its performance characteristics were determined by Mayo Clinic in a manner consistent with CLIA requirements. This test has not been cleared or approved by the US Food and Drug Administration.

CPT Code Information

88184

LOINC Code Information

Test ID Test Order Name Order LOINC Value
CD40 CD40 by Flow, QL, B 104694-5

 

Result ID Test Result Name Result LOINC Value
89009 CD40 by Flow, QL, B 104694-5

Interpretation

This assay is qualitative; CD40 expression is reported as present (normal) or absent (abnormal). Normal B cells express surface CD40 on the majority of cells.

 

Hyper-IgM (HIGM3) syndrome patients typically do not express CD40 on the surface of B cells. Genotyping of CD40 is required for a definite diagnosis of HIGM3. Call 800-533-1710 for ordering assistance.

Clinical Reference

1. Bishop GA, Hostager BS. The CD40-CD154 interaction in B cell-T cell liaisons. Cytokine Growth Factor Rev. 2003;14(3-4):297-309

2. Lee WI, Torgerson TR, Schumacher MJ, et al. Molecular analysis of a large cohort of patients with hyper immunoglobulin M (IgM) syndrome. Blood. 2005;105(5):1881-1890

3. Kutukculer N, Moratto D, Aydinok Y, et al. Disseminated cryptosporidium infection in an infant with hyper-IgM syndrome caused by CD40 deficiency. J Pediatr. 2003;142(2):194-196

4.Banday AZ, Nisar R, Patra PK et al. Clinical and immunological features, genetic variants, and outcomes of patients with CD40 Deficiency J Clin Immunol. 2023;44(1):17. doi:10.1007/s10875-023-01633-1.

5. Yazdani R, Fekrvand S, Shahkarami S, et al. The hyper IgM syndromes: Epidemiology, pathogenesis, clinical manifestations, diagnosis and management. Clin Immunol. 2019;198:19-30. doi:10.1016/j.clim.2018.11.007

Report Available

3 to 4 days

Method Name

Flow Cytometry