Test Code ASOU Arsenic Occupational Exposure, Random, Urine
Specimen Required
Only orderable as part of profile. For more information see:
-ASUOE / Arsenic Occupational Exposure with Reflex, Random, Urine
-HMUOE / Heavy Metal Occupational Exposure, with Reflex, Random, Urine
Useful For
Screening test for detection of occupational exposure to arsenic in random urine specimens
Testing Algorithm
If total arsenic concentration is 10 mcg/L or greater, then speciation will be performed at an additional charge.
Method Name
Only orderable as part of profile. For more information see:
-ASUOE / Arsenic Occupational Exposure with Reflex, Random, Urine
-HMUOE / Heavy Metal Occupational Exposure, with Reflex, Random, Urine
Triple-Quadrupole Inductively Coupled Plasma Mass Spectrometry (ICP-MS/MS)
Reporting Name
Arsenic Occupational ExposureSpecimen Type
UrineSpecimen Minimum Volume
3 mL
Specimen Stability Information
Specimen Type | Temperature | Time | Special Container |
---|---|---|---|
Urine | Refrigerated (preferred) | 28 days | |
Frozen | 28 days | ||
Ambient | 72 hours |
Reference Values
Only orderable as part of profile. For more information see:
-ASUOE / Arsenic Occupational Exposure with Reflex, Random, Urine
-HMUOE / Heavy Metal Occupational Exposure, with Reflex, Random, Urine
Biological Exposure Indices (BEI): <35 mcg/L at end of work week
Interpretation
Mayo Clinic uses the American Conference of Governmental Industrial Hygienists biological exposure index (BEI) as the reference value. The BEI is the sum of all the toxic species (inorganic arsenic plus methylated arsenic metabolites).
Physiologically, arsenic exists in a number of toxic and nontoxic forms. The total arsenic concentration reflects all the arsenic present in the sample regardless of species (eg, inorganic vs. methylated vs. organic arsenic). The measurement of urinary total arsenic levels is generally accepted as the most reliable indicator of recent arsenic exposure. However, if the total urine arsenic concentration is elevated, arsenic speciation must be performed to identify if it is a toxic form (eg, inorganic and methylated arsenic forms) or a relatively nontoxic organic form (eg, arsenobetaine and arsenocholine).
The inorganic toxic forms of arsenic (eg, As[III] and As[V]) are found in the urine shortly after ingestion, whereas the less toxic methylated forms, monomethylarsonic acid (MMA) and dimethylarsinic acid (DMA) are the species that predominate longer than 24 hours after ingestion. In general, urinary As(III) and As(V) concentrations peak in the urine at approximately 10 hours and return to normal 20 to 30 hours after ingestion. Urinary MMA and DMA concentrations normally peak at approximately 40 to 60 hours and return to baseline 6 to 20 days after ingestion.
This test can determine if a patient has been exposed to above-average levels of arsenic. It cannot predict whether the arsenic levels in their body will affect their health.
Clinical Reference
1. Fillol CC, Dor F, Labat L, et al. Urinary arsenic concentrations and speciation in residents living in an area with naturally contaminated soils. Sci Total Environ. 2010;408(5):1190-1194
2. Caldwell KL, Jones RL, Verdon CP, Jarrett JM, Caudill SP, Osterloh JD. Levels of urinary total and speciated arsenic in the US population: National Health and Nutrition Examination Survey 2003-2004. J Expo Sci Environ Epidemiol. 2009;19(1):59-68
3. Agency for Toxic Substances and Disease Registry: Toxicological profile for arsenic. US Department of Health and Human Services; August 2007. Available at www.atsdr.cdc.gov/ToxProfiles/tp2.pdf
4. Strathmann FG, Blum LM. Toxic elements. In: Rifai N, Chiu RWK, Young I, Burnham CD, Wittwer CT, eds. Tietz Textbook of Laboratory Medicine. 7th ed. Elsevier; 2023:chap 44
5. Keil DE, Berger-Ritchie J, McMillin GA. Testing for toxic elements: A focus on arsenic, cadmium, lead, and mercury. Lab Med. 2011;42(12):735-742 doi:10.1309/LMYKGU05BEPE7IAW
6. Navas-Acien A, Francesconi KA, Silbergeld EK, Guallar E. Seafood intake and urine concentrations of total arsenic, dimethylarsinate and arsenobetaine in the US population. Environ Res. 2011;111(1):110-118. doi:10.1016/j.envres.2010.10.009
7. Tchounwou PB, Yedjou CG, Udensi UK, et al. State of the science review of the health effects of inorganic arsenic: Perspectives for future research. Environ Toxicol. 2019;34(2):188-202. doi:10.1002/tox.22673
Performing Laboratory
Mayo Clinic Laboratories in RochesterTest Classification
This test was developed and its performance characteristics determined by Mayo Clinic in a manner consistent with CLIA requirements. It has not been cleared or approved by the US Food and Drug Administration.LOINC Code Information
Test ID | Test Order Name | Order LOINC Value |
---|---|---|
ASOU | Arsenic Occupational Exposure | 13463-5 |
Result ID | Test Result Name | Result LOINC Value |
---|---|---|
608890 | Arsenic Occupational Exposure | 13463-5 |
608891 | Total Arsenic Concentration | 5586-3 |
Day(s) Performed
Monday through Friday
Report Available
2 to 4 daysCPT Code Information
82175